Metadata

Field Value

Data Collector

Barber, David

Deployment ID

GL_LWH_M_2018

Research Area

manitoba-great-lakes

Deployment Dates

Deployment Dates 1

Deployment start date

2018-05-16

Deployment end date

2018-10-08

Activity

Deployed with ecotriplet 1442 and CT 1589

Notes

Location or Station ID

GL_LWH_M

Instrument Details

Instrument Details 1

Instrument Name

Ecotriplet 1442

Standardized Instrument Name

WETLabs ECO Triplet sensor

Instrument Unique ID

BBFL2WB1442

Instrument DOI

DOI

Description

A three-optical-sensor, user-defined instrument that may carry any combination of single-wavelength fluorometers or scattering meters. The fluorometers can be configured (with typical sensitivities) for chlorophyll (470/695 nm, sensitivity 0.015-0.025 ug/l), FDOM (370/460 nm, sensitivity 0.184 ppb), phycocyanin (630/680 nm, 0.086 ppb), phycoerythrin (520/595 nm, sensitivity 0.086 ppb), uranine (470/530 nm, sensitivity 0.073 ppb) or rhodamine (520/595 nm, 0.086 ppb). The scattering meter can typically measure optical scattering at blue, green or red wavelengths (412, 470, 532, 650, 700 and 880 nm, sensitivity 0.002-0.003 m-1)

Instrument Details 2

Instrument Name

Alec CT 1589

Standardized Instrument Name

Alec CT (conductivity, temperature) sampler

Instrument Unique ID

1589

Instrument DOI

DOI

Description

Conductivity and Temperature sensor used on Manitoba Great Lakes Moorings

Related Files